Model Number 357500
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2021
Event Type  malfunction  
Event Description
A passeo-18 balloon catheter was chosen for treatment.The balloon was placed properly inside the lesion.As the balloon was inflated up to 6 bar the pressure of the balloon would slowly turn down.To keep the balloon inflated at 6 bar it was necessary to keep on giving more pressure.
 
Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 54 mm.In close vicinity of the tear scratches were observed which have likely been caused by a hard, sharp-edged object such as e.G.Anatomical structure.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test and a pressure test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause is most likely related to the patients anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 6/60/130
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12540767
MDR Text Key273589238
Report Number1028232-2021-05348
Device Sequence Number1
Product Code LIT
UDI-Device Identifier07640119559117
UDI-Public07640119559117
Combination Product (y/n)N
PMA/PMN Number
K072765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number357500
Device Catalogue NumberSEE MODEL NO.
Device Lot Number08202341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Initial Date Manufacturer Received 09/27/2021
Initial Date FDA Received09/28/2021
Supplement Dates Manufacturer Received10/11/2021
Supplement Dates FDA Received10/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-