Model Number AS480
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/03/2021
Event Type  malfunction  
Manufacturer Narrative
The stopcock and syringe malfunction or leakage may potentially impact staining.No delay to testing was reported.No patient or user harm was indicated.Patient information has not been provided by the user.
 
Event Description
Customer reported false negative stainings.The field service engineer replaced the syringe and stopcock which were attributed to the event.Instrument was fully operational and returned to service.No indication of patient or user harm.No delay to testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE INTL. PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN  768923
Manufacturer (Section G)
SHANDON DIAGNOSTICS LIMITED
tudor road, manor park
runcom, cheshire WA7 1 TA
UK   WA7 1TA
Manufacturer Contact
bryce portier
1834 state highway 71 west
cedar creek, TX 78612
4082185467
MDR Report Key12546937
MDR Text Key273784497
Report Number3003423869-2021-00276
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-