• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC INQUIRY¿ OPTIMA¿ DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC INQUIRY¿ OPTIMA¿ DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81600
Device Problems Device Dislodged or Dislocated (2923); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2021
Event Type  malfunction  
Event Description
During an atrial fibrillation procedure, electrode 13 was displaced.The catheter was replaced and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
One duodecapolar, spiral loop, inquiry optima diagnostic catheter was received for evaluation.Electrode 13 was not displaced as initially reported.Dissection revealed conductor wire 13 had been fractured proximal to the weld joint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the fractured conductor wire 13 remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ OPTIMA¿ DIAGNOSTIC CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
2375 morse ave
irvine CA 92614
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12546946
MDR Text Key273784480
Report Number2030404-2021-00065
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K042775
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberIBI-81600
Device Lot Number7834222
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-