BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564840
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2021 that an ultraflex tracheobronchial covered distal release stent was to be implanted in the left lung to treat a malignant left middle segment bronchial stenosis during a bronchoscopy procedure performed on (b)(6) 2021.Reportedly, the patient's anatomy was not dilated prior to stent placement.During the procedure, the black stent deployment suture fractured and the stent was unable to fully deploy.The stent was partially deployed on the delivery system when it was removed from the patient.The procedure was not completed because another of the same device was unavailable.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key12547216
MDR Text Key273793875
Report Number3005099803-2021-05101
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2021
Device Model NumberM00564840
Device Catalogue Number6484
Device Lot Number0024420319
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/10/2021
Initial Date FDA Received09/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient Weight80
-
-