Model Number 45007 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: the event date was not provided and estimated based on the aware date.
|
|
Event Description
|
It was reported that the blades stopped spinning.A 2.4mm jetstream atherectomy catheter was selected for use.The blades stopped spinning during the procedure.However, infusion still functioned.The device was removed from the patient.
|
|
Manufacturer Narrative
|
B3 - date of event: the event date was not provided and estimated based on the aware date.
|
|
Event Description
|
It was reported that the blades stopped spinning.A 2.4mm jetstream atherectomy catheter was selected for use.The blades stopped spinning during the procedure.However, infusion still functioned.The device was removed from the patient.It was further reported target lesion was located in the common femoral artery.The physician attempted to restart the device several times during the procedure and outside the patient.The procedure was successfully completed with a new device.
|
|
Event Description
|
It was reported that the blades stopped spinning.A 2.4mm jetstream atherectomy catheter was selected for use.The blades stopped spinning during the procedure.However, infusion still functioned.The device was removed from the patient.It was further reported target lesion was located in the common femoral artery.The physician attempted to restart the device several times during the procedure and outside the patient.The procedure was successfully completed with a new device.
|
|
Manufacturer Narrative
|
B3 - date of event: the event date was not provided and estimated based on the aware date.Device eval by mfr: the jetstream device xc-2.4 was received for analysis.The shaft and the remainder of the device was inspected for damage.Visual examination showed a kink/buckling located 1cm from the tip.The functionality of the device was checked by setting up the product per the ifu.The device primed as designed.The device was activated, and the blades did not spin as designed.The devices motor was heard; however, no tip rotation was noticed.The devices pod was opened to inspect for damage.It was noticed the pinion gear had slipped off the drive shaft.When the pinion gear slides off the shaft and does not contact the motor gear, rotation of the tip would stop.The gear was slid back on the shaft and the device was activated.The device ran and the blades did spin for 3 minutes and the testing was completed.The gear did not slide off after it was slid back on for testing.Device analysis determined the condition of the returned device was consistent with the reported information of blades not spinning.
|
|
Search Alerts/Recalls
|