• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE CARDIOFORM ASD OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE CARDIOFORM ASD OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER Back to Search Results
Model Number ASD37A
Device Problem Break (1069)
Patient Problem Obstruction/Occlusion (2422)
Event Date 07/22/2021
Event Type  Injury  
Manufacturer Narrative
The gore® cardioform asd occluder instructions for use states: adverse events associated with the use of the occluder may include but are not limited to: device embolization.
 
Event Description
It was reported to gore a 37mm gore® cardioform asd occluder was selected to treat an atrial septal defect.It was reported the retrieval cord broke for some reason (very little force on it, i usually make sure everything is coaxial and do not pull with too much force when pulling on the cord).The device embolized to the left atrium/left ventricle.The device was able to be snared out and a second 37mm gore® cardioform asd occluder was implanted.It was reported the patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE CARDIOFORM ASD OCCLUDER
Type of Device
TRANSCATHETER, SEPTAL OCCLUDER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
kathy titus
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key12553968
MDR Text Key274033813
Report Number2017233-2021-02420
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00733132636501
UDI-Public00733132636501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Model NumberASD37A
Device Catalogue NumberASD37A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2021
Date Device Manufactured12/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age11 YR
Patient Weight30
-
-