• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH CARDIOHELP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CARDIOHELP-I
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/27/2021
Event Type  malfunction  
Manufacturer Narrative
A follow up will be submitted when additional information becomes available.A getinge service technician will investigate the cardiohelp in question.
 
Event Description
It was reported that the flow/bubble sensor quit reading.The event occured during treatment.The cardiohelp was changed.No harm to patient was reported.Complaint number: (b)(4).
 
Event Description
Complaint number: (b)(4).
 
Manufacturer Narrative
It was reported that the flow/bubble sensor (flow probe) quit reading during treatment.No harm to any person has been reported.A getinge technician was sent for investigation on 2021-10-01.He could not confirm the failure and the device was put back in use.The log files also shows no malfunction of the device.Therefore, the reported failure "flow/bubble sensor (flow probe) quit reading" could not be confirmed.An exact root cause could not be identified.However, with reference to the current risk file the following most possible causes could be linked to the reported failure: wrong bubble sensor information; influence due to other ultrasonic devices (e.G.Flow sensor); connection of non-capatible sensor; environmental influences (atmospheric pressure, temperature, humidity, emi, overvoltage); bubble sensor cannot be plugged.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key12559861
MDR Text Key274222657
Report Number8010762-2021-00534
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP-I
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-