• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2021
Event Type  malfunction  
Manufacturer Narrative
Part # 319.006, synthes lot # 4569369, supplier lot # n/a, release to warehouse date: 18 mar2003, manufactured by: synthes (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Visual inspection: the depth gauge f/scr ø2+2.4 meas-range up-t (p/n: 319.006, lot #: 4569369) was returned and received at us customer quality (cq).Upon visual inspection, it was observed that the needle component of the device was slightly bent.Also, the protection sleeve was not returned.Some light scratches were observed on the device surface, which could not impact the device functionality.No other issues were observed with the complaint device.Functional test: a functional test was performed.The needle component of the device didn't slide smoothly inside the body component of the device, it showed resistance.The bent condition of the needle component could have contributed to the device interaction issue.Can the complaint be replicated with the returned device(s)? yes.Dimensional inspection: measured dimensions: needle diameter = conforming.Document/specification review.Based on the date of manufacture, the current and manufactured revision of drawings were reviewed.Depth gauge for 2.0/ 2.4mm screws.Needle.Protection sleeve.No design issues or discrepancies were found during this investigation.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the needle component is bent and didn't slide smoothly inside the body component of the device.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.The potential cause for the bent needle could be due to unintended forces applied to the device and the potential cause for device interaction could be due to the bent condition of the needle.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on unknown date the patient underwent a unknown procedure.It was noted that two products were affected during procedure.The depth gauge was not sliding well.In the second portion, the surgeon had inserted a 2.4 va locking screw into the styloid of the distal radius plate.The screw did not lock into the plate and was just spinning.The surgeon chose a shorter screw to see if it would lock and it also did not lock into the plate.The surgeon ultimately removed the screw and placed a locking screw into a different hole.The surgery was completed successfully.There was a five minutes of surgical delay noted.No patient consequences noted.Concomitant device: unknown screws (part # unknown, lot # unknown, quantity # unknown).This report is for one (1) depth gauge for 2.0mm and 2.4mm screws.This is report 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12561511
MDR Text Key274250617
Report Number2939274-2021-05864
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier07611819707238
UDI-Public07611819707238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Device Lot Number4569369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN SCREWS.; VA-LCP-2-COLUMN DRP2.4 VOLAR LE SHAFT 3H.
Patient Age59 YR
Patient Weight74
-
-