• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENT HIP 53X28 BLU; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENT HIP 53X28 BLU; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Model Number 1035-53-000
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Fall (1848); Joint Dislocation (2374)
Event Date 09/17/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient had parkinson¿s and fell and dislocated his hip.The surgeon elected to length the ball to tighten the joint.Doi: (b)(6) 2021.Dor: (b)(6) 2021.Affected side: unknown hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 53X28 BLU
Type of Device
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12568079
MDR Text Key274456245
Report Number1818910-2021-21781
Device Sequence Number1
Product Code KWY
UDI-Device Identifier10603295003434
UDI-Public10603295003434
Combination Product (y/n)N
PMA/PMN Number
K812672
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1035-53-000
Device Catalogue Number103553000
Device Lot NumberHG5307
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/17/2021
Initial Date FDA Received10/04/2021
Supplement Dates Manufacturer Received10/28/2021
Supplement Dates FDA Received10/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARTICUL/EZE BALL 28 +5 BR.; SELF CENT HIP 53X28 BLU.; ARTICUL/EZE BALL 28 +5 BR; SELF CENT HIP 53X28 BLU
Patient Outcome(s) Required Intervention;
-
-