Catalog Number 201.928
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/04/2021
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, during a procedure the screw broke at the stem.Half of the screw remains in the patient.Another screw was used to complete the procedure.This report involves one (1) 2.0mm imf screw self-drilling 8mm.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: a product investigation was conducted.Visual inspection: the imf screw ø2 l8 sst (product code: 201.928, lot number: 145p649) was received at us cq.Upon inspection of the device it was noted that the threaded portion was broken and the broken piece was not returned.At the time of investigation there were no radiographs provided to confirm the embedded device condition.No other device defects were noted.Dimensional inspection: a dimensional inspection cannot be performed due to post manufacturing damage.Document/specification review: no design or manufacturing discrepancies were noted.Conclusion: the complaint condition is confirmed for the received device as part of the threaded portion of the imf screw broke off.The condition of embedded device cannot be confirmed as no evidence has been provided to support that allegation.A definitive assignable root cause cannot be determined from the provided information.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: product code: 201.928; lot #: 145p649; manufacturing site: mezzovico.Release to warehouse date: 11 may 2021.A manufacturing record evaluation was performed for the not sterile lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM IMF SCREW SELF-DRILLING 8MM
Type of Device
SCREW, FIXATION, INTRAOSSEOUS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12569138
MDR Text Key274522267
Report Number8030965-2021-08328
Device Sequence Number1
Product Code DZL
UDI-Device Identifier07611819724358
UDI-Public(01)07611819724358
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K010527
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201.928
Device Lot Number145P649
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/07/2021
Initial Date FDA Received10/04/2021
Supplement Dates Manufacturer Received10/13/2021
Supplement Dates FDA Received11/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/11/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-