Model Number N/A |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Cut (2587)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that the device does not work well, makes a weird/different sound and the cut graft is thicker than set.
|
|
Event Description
|
No additional information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).The investigation is complete.Review of the most recent repair record determined the device was out of calibration and control bar out of position and the bearings and plunger were replaced and device recalibrated and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|