Model Number N/A
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/02/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that the device does not work well, makes a weird/different sound and the cut graft is thicker than set.
 
Event Description
No additional information available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).The investigation is complete.Review of the most recent repair record determined the device was out of calibration and control bar out of position and the bearings and plunger were replaced and device recalibrated and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12576285
MDR Text Key274733642
Report Number0001526350-2021-01121
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number31788600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-