Model Number 865240
Device Problem No Audible Alarm (1019)
Patient Problem Cardiac Arrest (1762)
Event Date 10/01/2021
Event Type  Injury  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
The customer reported that a patient had a cardiac arrest and the intellivue mx800 patient monitor did not sound an alarm.
 
Manufacturer Narrative
Philips remote service engineer (rse) spoke to the customer.During the investigation, it was determined that the reported event was associated with a faulty intellivue multi measurement server x2 in use with the mx800.It was confirmed that there was no malfunction of the mx800 monitor.The engineer performed an evaluation/analysis and determined that the intellivue mx800 patient monitor was working as intended with no failures found.The reported event was caused by the intellivue multi measurement server x2 and is investigated in a separate complaint.
 
Event Description
The customer reported that a patient had a cardiac arrest and the intellivue mx800 patient monitor did not sound an alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX800 PATIENT MONITOR
Type of Device
INTELLIVUE MX800 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
derek sammarco
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key12582477
MDR Text Key274911453
Report Number9610816-2021-10399
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838020733
UDI-Public00884838020733
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865240
Device Catalogue Number865240
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight80 KG
-
-