• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW HEMOPRO 2; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW HEMOPRO 2
Device Problem Inflation Problem (1310)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Trackwise id # (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview hemopro 2.The pa doing the case had a hard time keeping the tunnel inflated.The case was going fine and the tunnel was inflating as he was doing the dissection with the scope.When they tried to switch over to the cutting tool/cannula, the tunnel wouldn't inflate.They tried several thing to troubleshoot.Everything they tried failed to keep co2 in the tunnel and inflating the tunnel.They went back to inserting the scope, and the tunnel did stay inflated.When using the scope is the only time the tunnel would stay inflated.Nothing else worked.The case was completed by opening a terumo device.No patient harm or incident.
 
Manufacturer Narrative
Trackwise (b)(4).Analysis of production for ship history conducted: (3331/213/67) a lot history record review was completed for lots 25159401, 25159402, and 25159584 the last 3 lots shipped to the account prior to the event/aware date.There were no ncmrs, rework, or deviations documented for the reported lot numbers.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period oct 2019 through sept 2021 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key12587238
MDR Text Key278625179
Report Number2242352-2021-00813
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00607567700406
UDI-Public00607567700406
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVASOVIEW HEMOPRO 2
Device Catalogue NumberVH-4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/14/2021
Initial Date FDA Received10/06/2021
Supplement Dates Manufacturer Received12/08/2021
Supplement Dates FDA Received12/08/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-