Model Number N/A |
Device Problems
Corroded (1131); Material Erosion (1214)
|
Patient Problems
Pain (1994); Metal Related Pathology (4530); Muscle/Tendon Damage (4532)
|
Event Date 07/27/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: item number 00620205622 item name shell porous with clusterholes 56 mm lot # 60548425; item number 00630505632 item name liner standard 32 mm i.D.For use with 56 mm o.D.Shell lot # 60918074; item number 00625006530 item name bone scr 6.5x30 self-tap lot # 60875832; item number 00625006525 item name bone scr 6.5x25 self-tap lot # 60740475.The device will not be returned for analysis location unknown; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2021 - 00310.
|
|
Event Description
|
It was reported by patient legal counsel initial right total hip arthroplasty preformed.Subsequently revised twelve (12) years post implantation due to pain.During the revision encountered a pseudocapsule, gross corrosion at the trunnion and acetabulum components, and gluteus medius damage from previous surgery.The head and liner were exchanged without complications.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Medical records were reviewed by an hcp and notes revision due to pain.Pseudocapsule in the joint and corrosion at the trunnion and screw holes of the shell.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2022 - 00348, 0002648920 - 2022 - 00022, 0002648920 - 2022 - 00023.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|