• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; IMAGINA GASTROSCOPE Back to Search Results
Model Number EG29-I10C
Device Problems Loss of or Failure to Bond (1068); No Display/Image (1183); Fluid/Blood Leak (1250); Moisture Damage (1405); Use of Device Problem (1670); Optical Distortion (3000)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2021
Event Type  malfunction  
Event Description
The images fades over, the electrical safety test failed, there is moisture inside the cmos chip.This event occurred at the time of before use.There was no report of patient harm.
 
Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model eg29-i10c-us is available in the usa with a 510k number k190805.If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Evaluation summary: the cmos chip and the distal body replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
IMAGINA GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12595416
MDR Text Key280997469
Report Number9610877-2021-01194
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04961333237586
UDI-Public04961333237586
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2021
Initial Date FDA Received10/07/2021
Supplement Dates Manufacturer Received09/16/2021
Supplement Dates FDA Received01/31/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/25/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-