Model Number V60
Device Problem Failure to Charge (1085)
Patient Problem Insufficient Information (4580)
Event Date 09/13/2021
Event Type  malfunction  
Event Description
It was reported that the ventilator was running on battery when it is connected to alternating current (ac) power and it was not charging.Although requested, it is unknown if the patient was connected to the ventilator at the time of the event.No patient or user harm was reported.The customer troubleshot with technical support and was advised to replace the power supply or power management board.The customer was provided with part numbers.Further information is pending.
 
Manufacturer Narrative
Multiple attempts were made to obtain additional information on the repair/service of the device that have been unsuccessful.Multiple attempts were made to obtain device context at the time of the reported failure; however, no further details were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key12598517
MDR Text Key275424437
Report Number2031642-2021-05229
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/13/2021
Initial Date FDA Received10/08/2021
Supplement Dates Manufacturer Received10/26/2021
Supplement Dates FDA Received11/22/2021
Date Device Manufactured10/21/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-