Model Number 74013480
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/29/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported during a tka procedure, that 2 nonrim speed pin 65 sterile broke off in a patients knee when the surgeon hit the screw in both the femur and tibia.The piece remains inside the patient, surgeon was not able to retrieve it.Procedure was completed with a backup device, no delay reported.An injury is reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the limited information provided, the clinical root cause of the reported breakage cannot be confirmed.The speed pin is an external communicating device composed of stainless steel and should have limited tissue/bone contact.However, since the broken piece remains in the patient¿s distal femur bone, micro-motion and/or migration is unlikely.The impact to the patient beyond that which has already been report cannot be determined.No further clinical assessment can be rendered at this time.Should any additional clinical documentation become available, the case may be re-evaluated.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This is a single use device, damage from misuse or rough handling are likely probable causes of the reported event.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONRIM SPEED PIN 65 STERILE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12611777
MDR Text Key275936689
Report Number1020279-2021-07524
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00885556070925
UDI-Public00885556070925
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74013480
Device Catalogue Number74013480
Device Lot Number20KNX0033
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/18/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexMale
-
-