• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EXPRT HIP; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 65MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EXPRT HIP; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 65MM Back to Search Results
Model Number 495-01-065
Device Problems Device Dislodged or Dislocated (2923); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 10/05/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - patient was dislocating so we swap head and proximal body for more stability.
 
Manufacturer Narrative
The reason for this revision surgery was reported as dislocation.The previous surgery and the surgery detailed in this event occurred 1 year and 1 month apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr#55852 associated with the main part #495-01-065, exprt revision hip, lateral offset proximal body, 65mm which documents that out of 15 parts lot, 4 parts were rejected due to major scratches at or near extraction hole.Later the rejected parts were reworked and accepted after retouched through spar.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to dislocation.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT HIP
Type of Device
EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 65MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key12618992
MDR Text Key276094800
Report Number1644408-2021-01101
Device Sequence Number1
Product Code KWZ
UDI-Device Identifier00190446153849
UDI-Public(01)00190446153849
Combination Product (y/n)N
PMA/PMN Number
K161610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number495-01-065
Device Catalogue Number495-01-065
Device Lot Number545U1105
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/05/2021
Initial Date FDA Received10/12/2021
Supplement Dates Manufacturer Received10/12/2021
Supplement Dates FDA Received10/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
400-03-282 LOT 858B1156
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age55 YR
-
-