• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-05
Device Problems Migration or Expulsion of Device (1395); No Audible Prompt/Feedback (2282); Migration (4003)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/14/2021
Event Type  Injury  
Event Description
The recipient is reportedly experiencing electrode migration and no sound.Imaging confirmed electrode migration.Revision surgery has been scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool marks on the antenna and the top cover.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical tests performed.The residual gas analysis test was compromised and data could not be collected.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
alexandra gonzalez
28515 westinghouse place
valencia, CA 91355
MDR Report Key12624215
MDR Text Key276370745
Report Number3006556115-2021-01502
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016878008
UDI-Public(01)07630016878008(11)210505(17)240430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/30/2024
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/13/2021
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received11/09/2021
04/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-