Model Number 83779
Device Problem Migration (4003)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/22/2021
Event Type  Injury  
Event Description
It was reported that the coil migrated, was snared but remained inside patient's body.The 90% stenosed target lesion was located in the severely tortuous and moderately calcified artery.A 5mm x 15cm interlock coil was selected for use.During the procedure, it was noted that the coil migrated to the superior mesenteric artery (sma).The physician attempted to snare the coil but was unsuccessful.The coil remained in the sma.The patient was then brought to other facility to finish the procedure.Computed tomography (ct) showed a flow in the sma post procedure.No further patient complications were reported and the patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Type of Device
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12624509
MDR Text Key276163373
Report Number2134265-2021-12447
Device Sequence Number1
Product Code KRD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number83779
Device Catalogue Number83779
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2021
Initial Date FDA Received10/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-