ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/24/2021
Event Type  malfunction  
Event Description
It was reported that during implant, the stylet was stuck in the left ventricular (lv) lead.When the physician attempted to remove the stylet, the terminal pin and inner lumen was detached from the lead, and the inner lumen was exposed.The physician elected to use a different lead to complete the procedure.The patient was discharged from the hospital after the procedure.
 
Manufacturer Narrative
The reported events of ¿inner lumen exposed, stylet was stuck in the lead, the terminal pin and inner lumen disengaged from the lead¿ were confirmed.As received, a complete lead was returned in two pieces with the stylet stuck inside the connector portion of the lead.Visual inspection found the connector pin and connector cap were pulled out of the connector assembly and the inner coil was stretched consistent with excessive forces during procedure.The ptfe coating of the stylet was found bunched up/clogged inside the inner coil distal to the connector pin.The cause of the reported events was due to bunched up ptfe coating of the stylet that prevented the removal of the stylet and excessive forces resulted in the connector pin and cap to be pulled out of the connector assembly.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12624807
MDR Text Key276174627
Report Number2017865-2021-32943
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2024
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000111741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/25/2021
Initial Date FDA Received10/13/2021
Supplement Dates Manufacturer Received12/03/2021
Supplement Dates FDA Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight83 KG
-
-