COVIDIEN VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM; STENT, ILIAC Back to Search Results
Model Number PXB35-05-17-080
Device Problems Device Dislodged or Dislocated (2923); Positioning Problem (3009)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 10/07/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use a visipro balloon expandable stent to treat a lesion in the common iliac artery.6fr sheath used.No embolic protection was used.Device prepped per ifu with no issue noted.It was reported that stent dislodgement occurred.It was reported that the initial stent (5mm) would not cross the intended lesion.When the delivery system was removed to dilate the artery it was noted the stent was no longer on the balloon.Using fluoroscopy they identified the stent in the iliac artery. the physician chose to use the second visipro stent to jail the dislodged stent in the iliac.The 7mm stent was used to pin the 5mm stent against the vessel wall.No alleged product issue with 7mm stent.The procedure was then completed without treating the intended lesion and with no patient harm.No further patient injury reported.
 
Manufacturer Narrative
Additional information: no vessel damage noted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM
Type of Device
STENT, ILIAC
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key12632556
MDR Text Key276398682
Report Number2183870-2021-00377
Device Sequence Number1
Product Code NIO
UDI-Device Identifier00643169787575
UDI-Public00643169787575
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/21/2023
Device Model NumberPXB35-05-17-080
Device Catalogue NumberPXB35-05-17-080
Device Lot NumberB073839
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2022
Date Device Manufactured08/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-