This event was reported by the distributor.The physician is: (b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
It was reported to boston scientific corporation that an endovive one step button was used during a percutaneous endoscopic gastrostomy procedure on (b)(6) 2021.During procedure, when the catheter was being pulled through the abdominal wall out of the body, the folded encased wings of the one step button device exited the abdominal wall.The bolster then detached and stayed between the stomach wall and the abdominal wall.The physician retrieved the bolster.The procedure was completed with a new endovive one step button.There were no patient complications reported as a result of this event.
|