The mayfield triad skull clamp (a1108) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: evaluation identified that the lock having both rotational and lateral movement and a residue buildup is present.Upon disassembly, it was observed that the index knob and the lock needed new components added to replace worn internal parts.Unit was received without the triad adjustment wrench.Unit was machined to have helicoils added to the large starburst threads, and new components were added to replace worn internal parts.Root cause: complaint confirmed via inspection of the unit.The index knob is unable to fully go into the locked position, and the index knob and lock required replacements for worn internal parts.Probable root cause is wear and tear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|