• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD TRIAD SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD TRIAD SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1108
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the mayfield triad skull clamp (a1108) would not lock or unlock.They added a lubricant but still would not lock or unlock.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
The mayfield triad skull clamp (a1108) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: evaluation identified that the lock having both rotational and lateral movement and a residue buildup is present.Upon disassembly, it was observed that the index knob and the lock needed new components added to replace worn internal parts.Unit was received without the triad adjustment wrench.Unit was machined to have helicoils added to the large starburst threads, and new components were added to replace worn internal parts.Root cause: complaint confirmed via inspection of the unit.The index knob is unable to fully go into the locked position, and the index knob and lock required replacements for worn internal parts.Probable root cause is wear and tear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD TRIAD SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
MDR Report Key12636326
MDR Text Key283817683
Report Number3004608878-2021-00601
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253518
UDI-Public10381780253518
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial
Report Date 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1108
Device Catalogue NumberA1108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Date Manufacturer Received09/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-