• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279401200
Device Problem Unintended Electrical Shock (4018)
Patient Problems Burn(s) (1757); Electric Shock (2554); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2021
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that the patient was shocked and potentially burned.
 
Manufacturer Narrative
Alleged failure: there was an electric shock which caused the patient to twitch.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be the tip comes in contact with a metal object during activation to create a short (sparks).The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that the patient was shocked and potentially burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12636522
MDR Text Key276557204
Report Number0002936485-2021-00547
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0279401200
Device Catalogue Number0279401200
Device Lot Number21089AE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/17/2021
Initial Date FDA Received10/14/2021
Supplement Dates Manufacturer Received09/17/2021
Supplement Dates FDA Received01/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-