Qn#(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.The manufacturer reports "in current manufacturing procedure, 100% leak testing and drop test are conducted at assembly area.Thus, any ineffective products will be culled out during this process.Therefore, it is very unlikely condition at the manufacturing area that the product having broken product released for shipment.As simulation, (b)(4) pieces of the same catalog number in current production were taken to test if there is any defect.Result shows no defect found on (b)(4) pieces of visual inspected samples.All the (b)(4) pieces sample passed leak test and drop test." without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|