• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE Back to Search Results
Model Number EG-3270UK
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
We checked the returned unit and confirmed that the ccd module defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module; however, other failure is not related to the alleged complaint.Based on the technical report, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
LINEAR ULTRASOUND VIDEO GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
MDR Report Key12654034
MDR Text Key281982525
Report Number9610877-2021-50098
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
PMA/PMN Number
K182004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial
Report Date 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG-3270UK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Initial Date Manufacturer Received 10/08/2021
Initial Date FDA Received10/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-