Catalog Number 361546 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 09/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: instrument sedi 40 17-42018 was returned to the manufacturer for service with respect to the reported defect ¿ noisy fan and broken tube cover.The instrument was evaluated by visual examination and functional testing and the fan was found to be noisy and a broken plate was found in the tube cover.A new cooling fan was installed and the plastic strip in the tube cover was replaced.Upon further testing the device passed all quality control criteria.
|
|
Event Description
|
It was reported when using the bd sedi-40 there was hardware/software malfunction for esr instrument.The following information was provided by the initial reporter translated to english.The customer stated: "the fan is doing strange noises.Issues with the mixing system - for example when instrument is powered on, the mixer starts automatically before to introduce the sample or do any procedure on the instrument.".
|
|