Model Number SC1200
Device Problem Decrease in Pressure (1490)
Patient Problem Rash (2033)
Event Date 07/30/2020
Event Type  Injury  
Manufacturer Narrative
File a 30-day mdr.An assessment was conducted.The event is still considered reportable under fda's regulation.Soclean has reviewed and made improvements to the complaint handling procedure to ensure compliance with fda regulations.These improvements were recommended because of an internal audit conducted by an external independent regulatory consultant.Soclean opened several capas to remediate the audit findings.Processes were revised to ensure reportable events were correctly identified and submitted to fda.Per fdas guidance, soclean is now reporting these events, which also includes retrospective items.
 
Event Description
Customer reports a rash with md intervention requiring clindamycin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCLEAN 2
Type of Device
SLEEP EQUIPMENT MAINTENANCE SYSTEM
Manufacturer (Section D)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer (Section G)
SOCLEAN, INC.
12 vose farm road
peterborough NH 03458
Manufacturer Contact
anne nadeau
12 vose farm road
peterborough, NH 03458
6033712533
MDR Report Key12661844
MDR Text Key277352258
Report Number3009534409-2021-00379
Device Sequence Number1
Product Code LRJ
UDI-Device Identifier00187293000860
UDI-Public187293000860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberSC1200
Device Catalogue NumberSC1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2021
Initial Date Manufacturer Received 07/30/2020
Initial Date FDA Received10/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-