MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU Back to Search Results
Model Number DM0010FAA
Device Problems Device Remains Activated (1525); Dull, Blunt (2407)
Patient Problems Hemorrhage/Bleeding (1888); Perforation (2001); Unintended Radiation Exposure (4565)
Event Date 09/28/2021
Event Type  Injury  
Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.This dm0010faa easydrill cranial perforator with lot number 836/20 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during deep brain stimulation procedure, the perforator did not stop while drilling for the second burr hole which is next to the coronal suture of the right hemisphere.It was reported that the dura mater of the patient was ruptured which requires hemostatic to control the bleeding and the ct scan revealed a presence of blood in the subarachnoid space.It was reported that an em200 and ad03 was used together with the perforator and it was noted that the perforator was not sharp as usual.It was also reported that there was a 10-minute procedure delay and the procedure was completed with the reported device.On follow-up, it was reported that the patient had no further issues post-surgery.It was also reported that the em200 and ad03 had no issue and will not be returned for evaluation.
 
Manufacturer Narrative
Report inconclusive.The device has been returned and is still pending their evaluation results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
On follow-up, it was reported that the ct scan procedure performed was part of the routine procedure.It was also reported that they can wait 1-2 days before doing the ct scan procedure, however, it was requested to be performed immediately after the surgery to check if there was an injury due to the event.
 
Manufacturer Narrative
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR  09980-39
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -39
BR   09980-39
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--39
BR   09980-39
140575722
MDR Report Key12669264
MDR Text Key278186574
Report Number1625507-2021-00333
Device Sequence Number1
Product Code HBF
UDI-Device Identifier07898959543197
UDI-Public07898959543197
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM0010FAA
Device Catalogue NumberDM0010FAA
Device Lot Number836/20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
EM200 (UNKNOWN S/N) AND AD03 (UNKNOWN S/N)
Patient Outcome(s) Other;
Patient Age30 YR
Patient SexMale
-
-