Model Number 0998-00-3023-53
Device Problem Suction Problem (2170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) called customer to set up repair date and time for service.However customer explained that their biomed engineer that performs their preventive maintenance service used a faulty balloon when performing pm and that they will no longer need our services.Fse is unable to clear this device for clinical use.A supplemental report will be submitted upon completion of our investigation.The first name of the initial reporter has been abbreviated due to field character limit; the full name should read (b)(6).The full name of the event site was shortened due to field character limit; the full name is (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) vacuum failed.It is unknown under which circumstances this event occurred; however there was no patient involvement, and no adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10, h11 corrected field: d1, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12675062
MDR Text Key281740891
Report Number2249723-2021-02418
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/01/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/30/2003
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-