An outside of the united states (ous) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered low and out of range for a single reagent pack well.The operator recalibrated the affected reagent pack well, after which qc recovered in range.Patient results were not reported on this pack only quality control samples were run.When the remaining tests in the affected pack well were exhausted, the system automatically moved to the second well of the reagent pack and also into a new reagent pack.Qc shifted high based on the calibration coefficients from the previous calibration using the affected pack well for both the second well of the sample pack and the new pack, indicating the second well of the same pack was not contaminated.The operator recalibrated using the new reagent pack and obtained acceptable qc results.No issues with other reagent pack wells were observed.The cause of the event is a contaminated reagent pack well.The reagent is performing according to specifications.No further evaluation of this device is required.
|
A discordant, falsely low total bilirubin_2 (tbil_2) result was obtained on a patient sample on an atellica ch 930 analyzer using atellica ch total bilirubin_2 reagent.The discordant result was not reported to the physician(s).The sample was repeated for tbil_2 on an alternate atellica ch 930 analyzer also using total bilirubin_2 reagent, recovering higher.The higher result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely low tbil_2 result.
|