Model Number 71992-01 |
Device Problem
High Readings (2459)
|
Patient Problems
Convulsion/Seizure (4406); Cramp(s) /Muscle Spasm(s) (4521)
|
Event Date 10/10/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high readings issue was reported with use of the adc freestyle libre 2 sensor.Customer received an unspecified high sensor scan result and experienced seizure, diarrhea and abdominal cramps.No treatment was required.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor 3mh005fpuch has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 5 (indicating normal termination).Inspected the plug assembly, no issues were observed.The current was applied to the sensor to perform linearity testing while in the test fixture.All results were within specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high readings issue was reported with use of the adc freestyle libre 2 sensor.Customer received an unspecified high sensor scan result and experienced seizure, diarrhea and abdominal cramps.No treatment was required.There was no report of death or permanent impairment associated with this event.
|
|