TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs, but was not able to reproduced error.While troubleshooting, fse observed functionality of the instrument while performing sample run and no damage was observed.Fse lubricated the hybrid arm tracks and sample nozzle spring, cleaned sample nozzle and cup pickup fingers and block, performed z-axis home position check, performed all alignments and found that the sample nozzle detach needed significant adjustment, fse adjusted the sample nozzle detach and proactively adjusted the rtc lane x.Fse validated the instrument by performing quality control run, all results passed without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: [4253] hybrid arm z-axis home overrun.Cause : the home sensor activated improperly after movement of the hybrid arm z-axis.If retry fails, the measurement result will be flagged (mf flag).Solution : contact tosoh service center or local representatives.The most probable cause of the reported event is due to misaligned sample nozzle to tip detach.
 
Event Description
A customer reported getting error message "4253 hybrid arm z axis home overrun" on the aia-2000 instrument.Instrument is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for luteinizing hormone (lhii), intact parathyroid hormone (ipth) and follicle stimulating hormone (fsh).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12678126
MDR Text Key277885417
Report Number3004529019-2021-00057
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/11/2021
Initial Date FDA Received10/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse