This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.1 device was functionally/visually inspected in the field.The device was repaired and returned to use.1 device was not evaluated, as the issue was identified and resolved during a troubleshooting call between the customer and stryker technical support.1 device was not evaluated and no cause was determined, as the customer did not make the device accessible for testing.There was no remedial action taken.This device is not labeled for single use.
|