DATEX-OHMEDA, INC. AVANCE; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Decrease in Suction (1146)
Patient Problem Insufficient Information (4580)
Event Date 09/28/2021
Event Type  malfunction  
Event Description
The hospital reported a malfunction resulting in weak suction.There was no report of patient involvement.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The coupling and manifold vacuum trap were replaced to correct the issue.Block a: no report of patient involvement.Block e1: the initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.H3 other text : a ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The coupling and manifold vacuum trap were replaced to correct the issue.Block a: no report of patient involvement.Block e1: the initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
elizabeth roedl
3030 ohmeda dr,
madison,, WI 53718
MDR Report Key12698982
MDR Text Key282850585
Report Number2112667-2021-02369
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received10/26/2021
Supplement Dates Manufacturer Received10/27/2021
Supplement Dates FDA Received11/24/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-