Daily quality controls were acceptable on both meters.The customer ran a correlation study and linearity studies on meter (b)(4) and the results were acceptable.Five strips of lot 479336 (vial (b)(4) roll (b)(4)) were received for investigation.The strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.There was no apparent reagent discoloration.More than 12 strips of lot 479336 (vial (b)(4) roll (b)(4)) were received for investigation.The strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.There was no apparent reagent discoloration.All testing with the returned strips produced acceptable results and no strip defects were observed.A routine retention testing process has been implemented.Valid retention results are available for all lots in the event a strip lot is alleged in a complaint.All retention data is reviewed on a monthly basis once testing is complete.
|
There was an allegation of questionable glucose results from accu-chek inform ii meters.At 7:05 am, the result from meter (b)(4) was 53 mg/dl from pic line.The staff thought his clinical presentation did not match this low blood sugar.At 7:08 am, the result from meter (b)(4) was 17 mg/dl from a fingerstick.At 7:10 am, the result from meter (b)(4) was 47 mg/dl from a fingerstick.At 7:11 am, the result from meter (b)(4) was 49 mg/dl from a fingerstick.At 7:13 am, the result from meter (b)(4) was 220 mg/dl.At 7:15 am, the result from meter (b)(4) was 188 mg/dl.At 7:25 am, the result from the laboratory was 38 mg/dl.
|