H10: only one (1) sample was received for evaluation.The other three (3) samples were not received and therefore, could not be evaluated.The sample that was received was a used segment together with one used sample of an extension tubing with a connector (non-baxter product).A visual inspection was performed to samples using the naked eye and it was observed that the male luer rigid component was cracked into the tubing connector of the extension set.Additionally, an appearance of white color was observed on the cracked components evidencing a trace of external force applied to the components.The reported condition was verified.The cause of the condition could not be determined, however, it was noted that the issue was possibly related to the force applied to disconnect the iv (intravenous) tubing from the extension tubing.Should additional relevant information become available, a supplemental report will be submitted.
|