MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BE-HLS 7050 #SHLS SET ADVANCED 7.0
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2021
Event Type  malfunction  
Manufacturer Narrative
Return of the affected product was requested but is still pending.A follow-up emdr will be submitted when additional information becomes available.
 
Event Description
The customer reported a blood leakage on the venous side of the oxygenator at the lower end.The hls set was replaced during treatment.
 
Manufacturer Narrative
It was reported that blood leaks on the venous side of the oxygenator at the lower end.The customer replaced the hls set during treatment.The hls set was investigated in the getinge laboratory on (b)(6) 2022.Hereby a damage of the luer lock connection of the sample line was confirmed which could led to a leak.The production records of the affected hls module were reviewed on (b)(6) 2022.According to the final test results, the affected hls module with udi#(b)(4) passed the tests as per specifications.Production related influences are unlikely to have contributed to the reported failure.The most probable cause of the reported failure was determined as an not tight connection between the sample line and the hls module which resulted in a leakage.Based on the investigation results the reported failure "leakage on the venous side of the oxygenator" could be confirmed.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLS SET ADVANCED
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key12705571
MDR Text Key278607592
Report Number8010762-2021-00576
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBE-HLS 7050 #SHLS SET ADVANCED 7.0
Device Catalogue Number701069073
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Reuse