DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is: (b)(6).(b)(6).(b)(4).
|
|
Event Description
|
It was reported that while servicing the cardiosave intra-aortic balloon pump (iabp) unit, the replacement fiber optic extension part was found defective during installation by a getinge field service engineer (fse).There was no patient involvement, and no adverse event reported.Related original complaint (b)(4).
|
|
Manufacturer Narrative
|
Additional contact information: name: (b)(6), e-mail: (b)(6), occupation: biomedical engineer.A getinge field service engineer (fse) evaluated the iabp unit and found that the fo sensor extension connector was improperly assembled.To fix the issue, the fse replaced the fo sensor extension connector, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
|
|
Event Description
|
N/a.
|
|