The reported complaint of the autopulse platform (sn (b)(4) does not power on with a fully charged autopulse li-ion battery, issue occurs intermittently was not confirmed during functional testing.The autopulse passed the initial functional test without any fault or error.The autopulse performed as intended.Upon visual inspection of the autopulse platform, unrelated to the reported complaint, a cracked front cover at the front end area was observed.The observed physical damage was likely due to mishandling, such as a drop.The front cover needs to be replaced to address this issue.During archive data review, no significant discrepancies were found.During service evaluation, the device was powered on and off multiple times without any issue.A short run_in test was performed, using the regular manikin with good known test batteries until discharged without any fault or error.The autopulse has passed all the testing and meets all required specifications.Multiple run_in and functional tests.
|