• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-6801A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-6801A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-6801A
Device Problems Power Problem (3010); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2021
Event Type  malfunction  
Event Description
The customer reported that the central nurse's station (cns) is having spontaneous shut down on its own.There was no patient injury reported.
 
Manufacturer Narrative
Technical support (ts) suggested for the customer to isolate the cns temporarily to a different power soruce as it could be that the power is surging and too much items are connected to the same uninterruptible power source (ups).The customer will try this and reach out to ts if needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6801A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
MDR Report Key12715419
MDR Text Key283267517
Report Number2080783-2021-01878
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921131640
UDI-Public04931921131640
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial
Report Date 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6801A
Device Catalogue NumberPU-681RA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/28/2021
Distributor Facility Aware Date10/01/2021
Device Age35 MO
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-