Catalog Number 875105
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Type  Injury  
Manufacturer Narrative
Therefore, because medical intervention was required, this event is reportable per 21 cfr part 803.Hp didn't get hot during testing.Visible black mark on outside of cap.Cap sticking inwards position coming intact with rotor.Black grime built up on outside of cap.Replaced cap and turbine.Replaced water hoses and water valve.
 
Event Description
It was reported that a midwest e pro handpiece overheated and burned a patient.The doctor administered antibiotics to treat the burn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PRO 1:5 CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key12723616
MDR Text Key279285909
Report Number9614977-2021-00052
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number875105
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/30/2021
Initial Date FDA Received10/29/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-