Model Number 173016 |
Device Problems
Component or Accessory Incompatibility (2897); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a vaginal cuff closure in a laparoscopic hysterectomy, the surgeon held the handle in the closed position while she toggled the device.Upon releasing the squeeze on the handle, the device opened and the needle fell out of the device and was stuck in the tissue.It was also noted that the device was difficult to toggle and load.To remove the needle, the surgeon had to pull needle from tissue.A new handle and needle was used to complete the case.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the needle fell out of the jaws of the device.Also, the device was difficult to load, and to toggle.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a vaginal cuff closure in a laparoscopic hysterectomy, the surgeon held the handle in the closed p osition while she toggled the device.Upon releasing the squeeze on the handle, the device opened and the needle fell out of the device and was stuck in the tissue.It was also noted that the device was difficult to toggle and load.To remove the needle, the surgeon had to pull the needle from tissue using a needle driver, the needle was still connected to the suture.A new handle and needle was used to complete the case.
|
|