INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD COMPOSITE SERIES SKULL CLAMP; ¿COMPOSITE SERIES¿ Back to Search Results
Model Number A3059
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the unit confirmed the complaint.The unit received had movement in the rocker arm assembly.General maintenance and cleaning performed and both the disk ratchet and retainer were replaced.Root cause - the reported complaint was confirmed.The unit received had movement in the rocker arm assembly and required replacement of the disk ratchet and retainer.Probable cause is wear and tear of the device.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
A facility reported that mayfield composite series skull clamp (a3059) has rotation and play (movement) that was present when the pins were locked.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12733896
MDR Text Key279627449
Report Number3004608878-2021-00618
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2021
Date Manufacturer Received10/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-