MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ8 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT DUOFIX TAP SZ8 HI OFF; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157012150
Device Problems Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cyst(s) (1800); Foreign Body Reaction (1868); Bone Fracture(s) (1870); Hematoma (1884); Hypersensitivity/Allergic reaction (1907); Muscular Rigidity (1968); Necrosis (1971); Pain (1994); Distress (2329); Discomfort (2330); Deformity/ Disfigurement (2360); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 06/21/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.
 
Event Description
Asr litigation record records.Records alleges severe pain, discomfort, injury, increased metal levels of cobalt and chromium, emotional distress, disability, disfigurement and economic damages.Doi: (b)(6) 2008 dor: (b)(6) 2021 left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: udi: (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
After review of medical records patient was revised was due to cystic pseudotumor, grossly debrided.All apparent metal appearing debris, significant corrosion present at the head and neck junction.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H6 health effect - clinical code: appropriate term / code not available (e2402) used to capture the medical device site joint movement impairment.
 
Event Description
In addition to what previously alleged pfs alleges pseudotumor, alval, femoral shaft perforation, fissure or fracture and lack of flexibility and mobility.
 
Event Description
Pfs received.In addition to what was previously alleged, pfs alleges metallosis, metal-tinged, necrotic appearing tissues, exuberant amount of black fluid and hematogenous tissue, massive bone loss and emotional distress.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, b6, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT DUOFIX TAP SZ8 HI OFF
Type of Device
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS 1818910
700 orthopaedic dr
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12738486
MDR Text Key279606827
Report Number1818910-2021-24174
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2017
Device Catalogue Number157012150
Device Lot NumberB52FM1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/20/2021
Initial Date FDA Received11/02/2021
Supplement Dates Manufacturer Received11/18/2021
11/30/2021
03/29/2022
04/22/2022
08/16/2022
Supplement Dates FDA Received11/24/2021
11/30/2021
03/31/2022
04/27/2022
08/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ADAPTER SLEEVES 12/14 +5; ASR ACETABULAR CUPS 58; ASR UNI FEMORAL IMPL SIZE 51; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL HEAD METAL ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient SexMale
Patient Weight113 KG
-
-