Catalog Number 365301
Device Problems Fluid/Blood Leak (1250); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: bd received fifty (50) samples and a photo for investigation.The samples and photo were reviewed and the customer¿s indicated failure mode for bowed tube was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode of bowed tube.Bd was not able to identify a root cause for the indicated failure mode.
 
Event Description
It was reported when using the bd vacutainer® sst¿ ii advance plus blood collection tubes there were tube extenders with molding defects that can be used.This event occurred 9 times.The following information was provided by the initial reporter.The customer stated: "the lab found distorted tubes.Distorted tubes were collected.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER SST II ADVANCE PLUS BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12739504
MDR Text Key279632684
Report Number9617032-2021-01112
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2022
Device Catalogue Number365301
Device Lot Number0349072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2021
Initial Date FDA Received11/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-