It was reported that the procedure was to treat a mildly calcified, mildly tortuous, 90% stenosed lesion in the internal carotid artery.A 9tx30mmx136 xact carotid stent system (sess) was advance to the target lesion and the actuator knob was rotated, however, the stent would not deploy.The actuator knob was rotated additional times until the stent became exposed but not expanded.The delivery system with the stent intact was removed and another xact stent was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
The device was returned for analysis.The reported activation/deployment failure was unable to be confirmed.It was noted that the sheath was not retracted and the stent was not exposed as reported.The deployment actuator was rotated to deploy the stent.The deployment actuator operated properly and the stent deployed with no anomalies noted.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that that the distal sheath of the delivery system was entrapped/bent and/or interaction with the mildly calcified, mildly tortuous and 90% stenosed anatomy resulted in the ratchet being unable to properly engage and release the stent from the delivery system; thus resulting in reported activation failure/ deployment failure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|