• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 100 - APAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 100 - APAC Back to Search Results
Model Number 27081
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The main circuit board was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device had a super capacitor leaking issue.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - APAC
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key12742676
MDR Text Key281139542
Report Number3004604967-2021-01118
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27081
Device Catalogue Number27081
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/04/2021
Date Manufacturer Received10/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-