According to the complaint the customer doubts the result of the lactate measurement on a abl90 flex analyzer after a spirometry.The measurement appears to be too low.There was no comparison measurement to back up the statement.Measurement deviation: measurement before spirometry: sample no.11094, measurement date & time: 18.10.2021, 12:32: lac 1,6 mmol/l (plausible).Measurement after spirometry (173 w load): sample no.11096, measurement date & time: 18.10.2021, 13:09: lac 2,1 mmol/l (implausible, customer expected a lactate value of 5,0 mmol/l).No reports of death or serious injury.
|
Radiometer investigation of the received datalogs has not identified a root cause.It is not possible to determine whether the device in question measures incorrectly or, how large the bias is, as there are no reference measurements available.Calibration and qc measurements for lactate does not reveals any malfunction for this sensor.
|